Replacing a 'Porosint' disc
Note: Figures and PDF version still required.
The Pore Pressure Cell Caps are fitted with sintered phosphor bronze filters. In time these 'Porosint' filters need replacing:-
-
Support the front face of the Pore Pressure Cell Cap and press out the existing filter disc from the back face. A piece of brass rod or similar, less then 10 mm in diameter, can be used as a drift either in a press or tapped lightly with a hammer.
-
Chamfer the new filter stone around one edge of its circumference to provide a lead-in. Press the stone into the Pore Pressure Cell Cap.
-
File the face of the new stone so that the face follows the curvature of the Cell Cap.
-
Drill and tap the centre of the stone I.S.O. metric coarse M2 x 0.4 mm pitch. Countersink at 90° the tapped hole on the front face so that the de-airing screw head sits level with the stone face.
-
As a result of the shaping of the stone (step [3] above) the surface will be smeared. With M2 screw fitted, very lightly grit blast the front face of the Pore Pressure Cell Cap to restore the porosity.
-
Alternatively, the cap can be acid etched.
-
Blow through the stone with compressed air.