Note: templates for the techniques discussed in this recipe are attached above. Both Microsoft Excel Macro Enabled Worksheet and LibreOffice Open Document Spreadsheet formats are available. These templates can be used to record a full calibration history for our instruments.
The Pore Pressure Cell Caps are fitted with sintered phosphor bronze filters. In time these 'Porosint' filters need replacing:-
Support the front face of the Pore Pressure Cell Cap and press out the existing filter disc from the back face. A piece of brass rod or similar, less then 10 mm in diameter, can be used as a drift either in a press or tapped lightly with a hammer.
The electronics units are sealed boxes fitted with desiccators. These absorb the small amount of moisture that, despite the sealing, eventually penetrates the units.
If the windows of the desiccators show pink then they must be changed. Spares are supplied, so fit the spare and dry the old desiccator.